2400 Series Wideband Spectrum Power Analyzer

$5,295$9,995

High-performance precision in both single and three-phases: a wattmeter, oscilloscope, and power spectrum analyzer in one. The 2400 power analyzer measures, computes and displays critical power variables so you can concentrate on more efficient & reliable testing.  The 2400 Series provides simultaneous, precise voltage and current measurements while monitoring and displaying the power parameters you need in the format that best fits your application.

Clear

2400 Options

+ $1,495

RS-232 Interface and Centronics printer output including Windows™ software to control, read, and store data

+ $1,595

RS-232- and IEEE-488 Interface, Centronics printer output including Windows™ software to control, read, and store data

+ $2,295

RS-232- and IEEE-488 Interface, Centronics printer output, and 4 analog outputs, 8 analog inputs including Windows™ software to control, read, and store data, read analog inputs

RS-232 and IEEE-488 Interface, Centronics printer output 4 analog outputs with provision to also output total power, 8 analog inputs including Windows™ software to control, read, and store data, read analog inputs

+ $1,095

1-Phase current sensor model 0-100A (supply by 2400)

+ $1,495

3-Phase current sensor model 0-100A (supply by 2400)

+ $325

2400 Series LabView driver

+ $325

2400 Series LabView-based standalone Software

+ $165

TTL input for external synchronization

+ $230

Network for artificial neutral

+ $1,395

0-300A, 0-3000A flexible current clamp (1 per Phase)

Product price
Additional options total:
Order total:

Precision Measurements for Any Waveform

Wattmeter, oscilloscope, and power spectrum analyzer

Extraordinary Features

  • The analyzer inputs are all galvanically isolated.
  • Broad band DC-300kHz.
  • Wide input range (0.3V – 1000V, 15mA – 40A).
  • Exceptional common mode rejection for use in frequency inverter driven systems.
  • The accuracy is 0.1% (0.05% versions are available).
  • The bright LCD monitor displays up to 10 measured values in well legible 9mm high numbers.
  • The Three-Phase Power Analyzer puts up to 32 measured values on the screen.

6 Configurations Available

1-Phase 1-Phase 1-Phase 3-Phase 3-Phase 3-Phase
No Harmonics Harmonics Analysis Harmonics Analysis No Harmonics Harmonics Analysis Harmonics Analysis
0.1% Accuracy 0.1% Accuracy 0.05% Accuracy 0.1% Accuracy 0.1% Accuracy 0.05% Accuracy
$5295 $5495 $5995 $7995 $8995 $9995
2410-1S 2410-1HS 2410-1HE 2430-3S 2430-3HS 2430-3HE

Calculated Values

Current/Voltage
  • RMS
  • RM
  • Peak
  • Harmonics
General
  • Frequency
  • Harmonic Distortion
  • Crest Factor
  • Form Factor
  • Oscilloscope Display
Integrated
  • Watt-Hours
  • VA-Hours
  • Amp-Hours
Power
  • Watts
  • VA
  • VAR
  • Power Factor

Variable Display

You have the choice to visualize wave forms, bar graphs or trend plots. A unique feature of these instruments allows a combination of meter mode and graphic modes.

Simple to Use

Checking power to determine the pertinent power parameters of a frequency inverter driven system is simple. All values display large letters easily read even in dark rooms. The user menu makes operation easy.

Export and Share Measurement Data

Send to PC

Send to Printer

Chart Recorder

[yith_wc_quick_order_form id=”1671″]

Voltage 8 ranges: 0.3 V, 1 V, 3 V, 10 V, 30 V, 100 V, 300 V, 1000 V
Frequency range DC, 0.1 Hz – 1 MHz
Crest Factor 3:1 at 50 % full scale (fs)
Input Impedance 1 MOhm
Common Mode 50 Hz/100 kHz 160 dB/100 dB
Standard accuracy 23°C; rms, mean, rectified mean; 0.3V typical

1 Hz-1 kHz                           ±(0.1 % rdg +0.1 % range)
DC, 1 kHz-10 kHz                     ±(0.2 % rdg +0.2 % range)
10 kHz-100 kHz                       ±(0.3 %/ range + 0.04 % /kHz rdg)
100 kHz-300 kHz                      ±(0.3 %/  range + 0.04 % /kHz rdg), typical

Improved accuracy
±(0.05 % rdg + 0.07 % range)
Current 13 ranges: 1.5 mA, 5 mA, 15 mA, 50 mA, 150 mA, 500 mA, 1.5 A, 5 A; 1, 3, 10, 30, 100 A Max. 1 A, 5 A, 30 A, resp.
Frequency range DC, 0.1 Hz-300 kHz / 1 MHz
Crest Factor 3:1 at 50 % full scale (fs)
Common Mode 50 Hz/100 kHz 160 dB/120 dB
Standard accuracy 23°C; 1 A-, 5 A-, shunt input               30 A input
1 Hz-1 kHz                ±(0.1 % rdg + 0.1 % rng)                  ±(0.1 % rdg + 0.1 % rng)DC, 1 kHz-10 kHz     ±(0.2 % rdg + 0.2 % rng)                  ±(0.7 % rdg + 0.2 % rng)10 kHz-100 kHz        ±(0.3 % range + 0.04 %/kHz rdg)     ±(0.3 % rng + 0.5 %/kHz rdg), typ100 kHz-300 kHz      ±(0.3 % range + 0.04 %/kHz rdg), typical
Lowest ranges 1.5 mA, 15 mA,
1 A: typical.
Improved accuracy 1Hz-400 Hz
±(0.05 % rdg + 0.07 % range)
Power

 

104 ranges corresponding to the products V x A.
Frequency range DC, 0.1 Hz-300 kHz
45 Hz-65 Hz
1 Hz-1 kHz
DC, 1 kHz-10 kHz 10 kHz-100 kHz(0.1 % rdg + 0.01 % range)
Add accuracy percentage figures of current and voltage,
+0.04 %/kHz PF
PF= 0 to ±0.1 PF= 0 to ±1 PF= 0 to ±1 PF=1
Frequency 0.1 Hz-400 kHz, V triggered; Accuracy ±0.1 %.
Computed Values Accuracy; Reactive Power, Var=±(VA2-W2)1/2, Apparent Power: VA=Arms Vrms; Power Factor: PF=W/VA; Crest Factor: CF=Ap/Arms, Vp/Vrms: Form Factor: FF=At/Arms, Vt/Vrms; Impedance: Z=Vrms/Arms; Total Harm Dist: THD=(Irms2- Ifund2)1/2/Irms

 

 

Add accuracy percentage figures of values involved in computation.
Integrator Energy, Charge; Accuracy Wh, Vah, Varh, Ah; Basic accuracy of integrated quantity.
Harmonic Analysis Frequency range of fundamental            2.5 Hz-100 kHz
Range of harmonic 1-99
Accuracy, Harmonic current and voltage
2 Hz-1 kHz              ±(0.1 % rdg + 0.1 % range) 1 kHz-10 kHz          ±(0.5 % rdg + 0.5 % range)
10 kHz.-100 kHz     ±(0.7 % range + 0.1 %/kHz rdg), typical
Display Blue liquid crystal graphic display with FL backlight               64×120 mm;  128 x 240 pixels

 

 

Power AC, 50-400 Hz; Fuse: Power

 

 

85 V-240 V; 2 A, 15 VA
Dieletric Strength Inputs to case or power supply Line input to case
Input to Input
2.5 kV/50 Hz/1 minute
1.5 kV/50 Hz/ 1 minute 4 kV/50 Hz/1 minute
Dimension H x W x D; Weight 150 x 235 x 320 mm; 4 kg
Options IEEE-488-2, RS232, Centronics printer output
4 programmable analog outputs; single-, sum-, or average values 4 analog inputs 0-±5V, input impedance 200 kΩ
4 analog inputs, 0-±10 V, input impedance 200 kΩ Rack Mounting Kit
Windows Operating Software 95, 98, 2000, NT, XP; transformer-motor testing
0-±5 V, accuracy 0.2 % 0-±5 V, accuracy 0.2 % 0-±10 V accuracy 0.2 %
1.5mA-1A Inp/ Shunt Input 1 A input Hi against ILo Shunt Hi                   Shunt Lo 1 A input,    mA:     1.5, 5, 15, 50, 150, 500, 1500
Shunt input, mV:    60, 60Ö10, 600, 600VÖ10, 6000, 6000Ö10
Input impedance:   60k
1 A input: set scaling to 0.1 Shunt input: 60 mV corresponds to 1.0000 A

 

Instrument Accessories

Maximize Your Instrument’s Performance with Certified-Compatible Cables, Leads, Connectors, & More

We recommend the accessories below for your instrument.  Each has been specifically designed and configured for top performance with our precision instruments.

Documents & Downloads

Here you can download the documentation relevant to your instrument. This section will be continually updated as more media is added (tutorials, demos, etc).

Manual and Specifications

What’s Included

All instrument orders include the following:

13-Month Comprehensive Instrument Warranty

All pre-paid orders include 13-month instrument warranty that covers component and labor costs in the event of needed repair or recalibration.

Certificate of Calibration

A fully-documented NIST-traceable Z540 certificate of calibration is included with your instrument.

Unlimited Phone & Email Support

Our engineering, sales, and customer support teams are always available to answer questions, walk you through particular product features, and assist with any issue that may arise.

Required Power Supply/Battery Pack

Unless you select a different option, every instrument ships with the cables and power supplies required to operate.

All Instrument Documentation

Includes user manual, calibration procedures, specifications, and any relevant materials.

Reviews

There are no reviews yet.

Be the first to review “2400 Series Wideband Spectrum Power Analyzer”

Your email address will not be published. Required fields are marked *

Quick Contact

Use this form to quickly configure and request a formal quote from our sales department. You may always reach us at 800-548-9806 if you'd like immediate assistance.
  • This field is for validation purposes and should be left unchanged.