Alpha 4314KRC Digital Igniter Tester

The Alpha 4314 Digital Igniter Tester is a portable digital ohmmeter which has been specifically designed for ultra-safe resistance testing on explosive devices (i.e. squibs, detonators).  

  • Intrinsically-safe circuitry
  • Standard test and fail-safe currents
  • 100 micro-ohm resolution up to 200M ohms
  • Configurable with 4 of 9 Ranges
  • NIST-traceable certificate of calibration included
  • Portable with rechargeable batteries
  • Now available with RS-232 interface
  • NSN 6625-01-460-1499

The Alpha 4314 KRC is a configuration of the Alpha 4314 that features reduced test currents.

Rng # Range Resolution Test Current Fail-Safe Current Accuracy
0 100µΩ 5mA 8mA ± 0.02% of reading ± 0.02% of range
1 20Ω 1mΩ 5mA 8mA ± 0.02% of reading ± 0.02% of range
2 200Ω 10mΩ 500μA 1.8mA ± 0.02% of reading ± 0.02% of range
3 2kΩ 100mΩ 50μA 180μA ± 0.02% of reading ± 0.02% of range
4 20kΩ 5μA 18μA ± 0.02% of reading ± 0.02% of range
5 200kΩ 10Ω 500nA 1.8μA ± 0.05% of reading ± 0.05% of range
6 2MΩ 100Ω 50nA 180nA ± 1% of reading ± 0.2% of range

Ranges, Resolution, and Accuracy

While standard version of the Alpha 4314 features 20, 200, 2k, and 20k Ohm ranges, it can be configured to meet your needs. The table below lists the nine possible ranges that can be installed in the Alpha 4314 KRC.  When ordering, simply specify the four range numbers you want to be installed in your instrument.  The 20MΩ and 100MΩ ranges are only available on the standard current configuration.

Alpha 4314 THIS PRODUCT

Alpha 4314 KRC Reduced Current

Rng # Range Resolution Test Current Fail-Safe Current Test Current Fail-Safe Current Accuracy
0 100µΩ 10mA 16mA 5mA 8mA ± 0.02% of reading ± 0.02% of range
1 20Ω 1mΩ 10mA 16mA 5mA 8mA ± 0.02% of reading ± 0.02% of range
2 200Ω 10mΩ 1mA 1.8mA 500μA 1.8mA ± 0.02% of reading ± 0.02% of range
3 2kΩ 100mΩ 100μA 180μA 50μA 180μA ± 0.02% of reading ± 0.02% of range
4 20kΩ 10μA 18μA 5μA 18μA ± 0.02% of reading ± 0.02% of range
5 200kΩ 10Ω 1μA 1.8μA 500nA 1.8μA ± 0.05% of reading ± 0.05% of range
6 2MΩ 100Ω 100nA 180nA 50nA 180nA ± 1% of reading ± 0.2% of range
7 20MΩ 1kΩ 10nA 18nA n/a n/a ± 2% of reading ± 0.2% of range
8 200MΩ 10kΩ 1nA 1.8nA n/a n/a ± 3% of reading ± 1% of range

TECHNICAL SPECIFICATIONS

Weight 3 lbs / 1.3kg net; 6lbs / 3kg shipping
Dimensions 9.5 × 11 × 3 in
Display Type 4 1/2 Digits LED
Overload Indication O.L.
Conversion Rate 3 Readings/Second
Terminal Configuration 4-Wire Kelvin
Battery Charger 9VDC at 300mA nominal
Current Source Compliance Clamped at 1.6V
Operating Temperature Range 0°C to 50°C
Power Requirements 4 “D” Cell 1.2V Rechargeable NiMH Batteries 10000mAh
Storage Temperature Range -10°C to 70°C
SKU: Alpha-4314KRC Categories: ,

Protective Fail-Safe Circuitry

Those working with explosive devices can have peace of mind with the Alpha 4314’s redundant fail-safe limiting circuitry that physically prohibits accidental ignition in the event of equipment failure.

9 Test Ranges

Select 4 of 9 test ranges when you order, giving you the ultimate in flexibility no matter what your application

4-Wire Kelvin Inputs

Never worry about lead wire length or contact resistance errors requiring redundant calculations with the Alpha 4314’s 4-wire Kelvin inputs.

Battery-Powered Portability

The Alpha 4314 is powered by high-capacity rechargeable batteries, enabling true portability and enhancing intrinsic safety. Continuous isolated AC power is available as an alternative (Option AC-115).

US Military-Approved

US Military-Approved
The Alpha 4314 has been approved by the Department of Defense and all branches of the armed services, including special projects use (i.e. Cruise Missle).

Export Data to PC/Printer/Recorder

You can configure your instrument with a BCD or RS-232 interface which enables the Alpha 4314 to communicate data with other devices.

 

Designed with Fail-Safe Circuitry

Countless critical military, aerospace, and industrial applications utilize squibs and other blasting agents, and it is vital that these devices detonate precisely when required – not sooner, not later, not while testing for continuity integrity.  To guarantee both the readiness of the device and the safety of the technician or engineer, the Alpha 4314 is equipped with redundant, dual fail-safe, current-limiting network circuitry that ensures instrument test currents cannot exceed the specified amount.

To certify that this level of safety is included in your instrument, we simulate a worst-case component failure during final testing and include the resulting recorded output current flow with your order.

Government and Military Approved

The 4314 series’ circuitry has been reviewed and approved by a U.S. Government Safety Board and all branches of our armed forces for specific military programs (i.e. Cruise Missile).

EWR Drop Test Compliant

Compliant To The Eastern And Western Range (Ewr) 127-1 Range Safety Requirements For Drop Test, including that “These devices shall be analyzed to verify that rough handling, dropping, or single component failure will not result in negating the current-limiting feature.”

Safe for Explosive and Volatile Atmospheres

The Alpha 4314 offers unmatched safety while delivering uncompromising performance and has passed rigorous testing to achieve its status as “Certified Mil Std 810 – Method 511 Safe for Explosive & Volatile Atmospheres.”

Select Options & Add to Cart

Range Configuration
Includes standard 20Ω 200Ω 2kΩ 20kΩ ranges

Includes standard 20Ω 200Ω 2kΩ 20kΩ ranges

Select 4 custom ranges to include below

Select 4 custom ranges to include below

Additional Rear Interface

Adds RS-232 Rear-Interface. Data Sheet Available

Adds RS-232 Rear-Interface.

Adds Binary Coded Decimal rear interface. Data Sheet Available

Adds Binary Coded Decimal rear interface.
Instrument Options

Replace battery functionality with isolated, continuous 115V AC power

Replace battery functionality with isolated, continuous 115V AC power

17025 Certification is performed on our behalf by a third party. Please contact us for additional information

17025 Certification is performed on our behalf by a third party. Please contact us for additional information
Recommended Accessories
HCC Protective Hard Carrying Case $605.00
R-4 Rack Mount Adapter $425.00
Instrument Warranty

Product price: $3,565.00
Total options: